• DocumentCode
    2706893
  • Title

    On the use of averaging for auditory evoked response detection

  • Author

    Sprevak, D. ; McAllister, H.G. ; McCullagh, P.J.

  • Author_Institution
    Fac. of Inf., Ulster Univ., Coleraine, UK
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    178
  • Lastpage
    183
  • Abstract
    Quantifies the relationship between the signal-to-noise ratio (SNR) and the number of realisations of the auditory brainstem response. The commonly-held belief that the SNR increases with the number of terms being averaged originates from the assumption that repeated realisations of the evoked potentials are perfectly correlated. We show that, in practice, the correlation is far from perfect, and this dramatically affects the number of observations which are required to obtain the desired SNR. The relationship reduces to the accepted form for perfectly correlated signals
  • Keywords
    auditory evoked potentials; brain; correlation theory; medical signal processing; noise; auditory brainstem response realizations; auditory evoked response detection; averaging; imperfect correlation; observation number; perfectly correlated signals;
  • fLanguage
    English
  • Publisher
    iet
  • Conference_Titel
    Advances in Medical Signal and Information Processing, 2000. First International Conference on (IEE Conf. Publ. No. 476)
  • Conference_Location
    Bristol
  • ISSN
    0537-9989
  • Print_ISBN
    0-85296-728-4
  • Type

    conf

  • DOI
    10.1049/cp:20000335
  • Filename
    889969