• DocumentCode
    270811
  • Title

    Electromagnetic tracker feasibility in the design of a dental superstructure for edentulous patients

  • Author

    Moreira, Antonio H. J. ; Queirós, Sandro ; Rodrigues, Nuno F. ; Pinho, A.C.M. ; Fonseca, Jaime C. ; Vilaça, João L.

  • Author_Institution
    Sch. of Health Sci., ICVS/3B´s - PT Gov. Associate Lab., Guimarães, Portugal
  • fYear
    2014
  • fDate
    11-12 June 2014
  • Firstpage
    1
  • Lastpage
    6
  • Abstract
    The success of the osseointegration concept and the Brånemark protocol is highly associated to the accuracy in the production of an implant-supported prosthesis. One of most critical steps for long-term success of these prosthesis is the accuracy obtained during the impression procedure, which is affected by factors such as the impression material, implant position, angulation and depth. This paper investigates the feasibility of 3D electromagnetic motion tracking systems as an acquisition method for modeling full-arch implant-supported prosthesis. To this extent, we propose an implant acquisition method at the patient mouth and a calibration procedure, based on a 3D electromagnetic tracker that obtains combined measurements of implant´s position and angulation, eliminating the use of any impression material. Three calibration algorithms (namely linear interpolation, higher-order polynomial and Hardy multiquadric) were tested to compensate for the electromagnetic tracker distortions introduced by the presence of nearby metals. Moreover, implants from different suppliers were also tested to study its impact on tracking accuracy. The calibration methodology and the algorithms employed proved to implement a suitable strategy for the evaluation of novel dental impression techniques. However, in the particular case of the evaluated electromagnetic tracking system, the order of magnitude of the obtained errors invalidates its use for the full-arch modeling of implant-supported prosthesis.
  • Keywords
    bone; calibration; dentistry; orthotics; prosthetics; Branemark protocol; calibration; dental superstructure; edentulous patients; electromagnetic motion tracking systems; electromagnetic tracker feasibility; full-arch implant-supported prosthesis; implant acquisition method; implant-supported prosthesis; osseointegration; tracking accuracy; Accuracy; Calibration; Electromagnetics; Implants; Interpolation; Polynomials; Prosthetics; distortion calibration; edentulous patients; electromagnetic tracker; implants measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Medical Measurements and Applications (MeMeA), 2014 IEEE International Symposium on
  • Conference_Location
    Lisboa
  • Print_ISBN
    978-1-4799-2920-7
  • Type

    conf

  • DOI
    10.1109/MeMeA.2014.6860046
  • Filename
    6860046