DocumentCode :
2709305
Title :
SAR Evaluation of Pregnant Woman Models in 64 MHz MRI Birdcage Coil
Author :
Shamsi, S. ; Wu, D.G. ; Chen, J. ; Liu, R. ; Kainz, W.
Author_Institution :
Dept. of Electr. & Comput. Eng., Houston Univ., TX
fYear :
2006
fDate :
11-16 June 2006
Firstpage :
225
Lastpage :
228
Abstract :
This study presents numerical SAR evaluation in pregnant woman models exposed to magnetic resonance imaging (MRI) radio frequency (RF) coil at 64 MHz. Nine pregnant woman models, representing different stage of pregnancy, are used for this study. SAR values in and around fetuses at different pregnancy stages are calculated. The investigations show an SAR increase beginning at the fifth month of pregnancy and local fetus SAR values above the IEC limit of 10 W/kg for the first level controlled mode. Based on our results we recommend not to perform MRI procedures for pregnant women in the first level controlled mode. These results can be used towards developing safety standards for pregnant woman undergoing an MRI
Keywords :
biological effects of fields; biomedical MRI; finite difference time-domain analysis; obstetrics; 64 MHz; FDTD simulations; IEC limit; birdcage coil; magnetic resonance imaging; pregnant woman model; safety standards; specific absorption rate; Animals; Biological system modeling; Coils; Finite difference methods; Magnetic resonance imaging; Pregnancy; Radio frequency; Resonant frequency; Safety; Time domain analysis; FDTD simulations; MRI RF coil; SAR; pregnant woman model;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Symposium Digest, 2006. IEEE MTT-S International
Conference_Location :
San Francisco, CA
ISSN :
0149-645X
Print_ISBN :
0-7803-9541-7
Electronic_ISBN :
0149-645X
Type :
conf
DOI :
10.1109/MWSYM.2006.249455
Filename :
4014865
Link To Document :
بازگشت