Title :
Development of planar microelectrode array for extracellular characterization of action potentials in sinoatrial node cells
Author :
Zhang, Xin ; Zhang, Yulong ; Li, Yingxin ; Nelson, Richard D. ; LaRue, John C.
Author_Institution :
Neural Eng. Res. Lab., Z&L Creative Corp., Lewes, DE, USA
Abstract :
In this research work we prototyped a planar microelectrode arrays that utilize SU-8 as the insulating material for non-invasive chronically characterization of action potentials from SA node cells. The microelectrode arrays are fabricated to form a microchannel based extracellular action potential characterization system which comprises a 2 by 50 linear microelectrode array for acquiring the action potentials with the principle of capacitive coupling and a full customized CMOS analog bio-signal readout ASIC for bio-signal conditioning and processing. The prototyped multi-channel mixed signal fully-customized CMOS biosensor analog front-end serves as the low magnitude action potential signal readout circuitry, which is comprised of analog signal buffers, high gain amplifier and control/interface. The concept of using epoxy-based photoresist SU-8 as the insulator instead of the conventional silicon nitride or silicon dioxide is due to the high insulation property and bio-compatibility of the SU-8 material. This biocompatible planar microelectrode arrays with microchannels provides the biological scientists powerful tools for the long term non-invasive study of action potentials from SA node cells to screen libraries of chemical compounds for new drug candidates.
Keywords :
bioelectric potentials; biomedical electronics; biomedical materials; biosensors; cardiology; cellular biophysics; microelectrodes; photoresists; readout electronics; SA node cells; action potentials; analog signal buffers; bio-signal conditioning; biocompatible planar microelectrode arrays; capacitive coupling; customized CMOS analog bio-signal readout ASIC; drug candidates; epoxy-based photoresist SU-8; extracellular characterization; high gain amplifier; insulating material; low magnitude action potential signal readout circuitry; microchannel based extracellular action potential characterization system; multi-channel mixed signal fully-customized CMOS biosensor analog front-end; planar microelectrode array; sinoatrial node cells; Arrays; Biosensors; Electric potential; Insulators; Microelectrodes;
Conference_Titel :
Bioelectronics and Bioinformatics (ISBB), 2011 International Symposium on
Conference_Location :
Suzhou
Print_ISBN :
978-1-4577-0076-7
DOI :
10.1109/ISBB.2011.6107630