• DocumentCode
    2715674
  • Title

    An analog beamformer for integrated high-frequency medical ultrasound imaging

  • Author

    Gurun, Gokce ; Zahorian, Jaime ; Tekes, Coskun ; Karaman, Mustafa ; Hasler, Paul ; Degertekin, F. Levent

  • Author_Institution
    Sch. of Electr. & Comput. Eng., Georgia Inst. of Technol., Atlanta, GA, USA
  • fYear
    2011
  • fDate
    10-12 Nov. 2011
  • Firstpage
    65
  • Lastpage
    68
  • Abstract
    We designed and fabricated a dynamic receive beamforming integrated circuit (IC) in 0.35-μm CMOS technology. This beamformer is suitable for integration with an ultrasound annular array for high-frequency (30-50 MHz) intravascular ultrasound (IVUS) imaging. The beamformer IC is capable of buffering, delaying and preamplification for 8 receive channels. We explored an analog delay cell based on a current-mode first-order all-pass filter, which is used as the basic building block to form an analog dynamic delay line. We also explored a bandwidth enhancement method on the delay cell that improved the overall bandwidth of the delay line by a factor of 6. Each delay cell consumes 2.1 mW of power and is capable of generating a tunable delay between 1.75 ns to 2.5 ns, enabling dynamic receive beamforming over a focal range from 1.4 mm to 2 mm. We successfully integrated the fabricated beamformer IC with an 8-element annular array. Our experimental test results demonstrated the desired buffering, preamplification and delaying capabilities of the beamformer.
  • Keywords
    CMOS analogue integrated circuits; all-pass filters; delays; high-frequency effects; medical image processing; ultrasonic imaging; 8-element annular array; CMOS technology; analog beamformer; analog delay cell; analog dynamic delay line; buffering; current-mode first-order all-pass filter; dynamic receive beamforming integrated circuit; frequency 30 MHz to 59 MHz; integrated high-frequency medical ultrasound imaging; power 2.1 mW; preamplification; receive channels; size 0.35 mum; time 1.75 ns to 2.25 ns; ultrasound annular array; Arrays; Bandwidth; Delay; Delay lines; Integrated circuits; Mirrors; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference (BioCAS), 2011 IEEE
  • Conference_Location
    San Diego, CA
  • Print_ISBN
    978-1-4577-1469-6
  • Type

    conf

  • DOI
    10.1109/BioCAS.2011.6107728
  • Filename
    6107728