• DocumentCode
    2716961
  • Title

    An 8-channel readout front-end for long-term sleep quality monitoring

  • Author

    Pu, Xiaofei ; Zhang, Hui ; Qin, Yajie ; Hong, Zhiliang

  • Author_Institution
    State Key Lab. of ASIC & Syst., Fudan Univ., Shanghai, China
  • fYear
    2011
  • fDate
    10-12 Nov. 2011
  • Firstpage
    385
  • Lastpage
    388
  • Abstract
    An 8-channel readout front-end (RFE) for long-term sleep quality monitoring is presented in this paper and features high common-mode rejection ratio (CMRR) and low input referred noise. Each channel is composed of an AC coupled instrumentation amplifier (IA) with chopping spike filter (CSF), a programmable gain amplifier (PGA), and a buffer, while the bias generator and non-overlapping clock are shared by all channels. The proposed circuit, built in standard 0.35 μ m CMOS technology, consumes 101 μ A from 2.7 V, while occupying 5 mm2 of chip area. According to the simulation, the AC coupled IA´s CMRR is 118 dB and input referred noise is merely 0.55 μ Vrms. Meanwhile, the RFE is digitally programmable for different applications.
  • Keywords
    CMOS integrated circuits; amplifiers; biomedical electronics; patient monitoring; readout electronics; sleep; 8-channel readout front-end; AC coupled instrumentation amplifier; IA CMRR; RFE; bias generator; buffer; chopping spike filter; current 101 muA; high common-mode rejection ratio; long-term sleep quality monitoring; low input referred noise; nonoverlapping clock; programmable gain amplifier; voltage 2.7 V; Choppers; Electrodes; Gain; Instruments; Monitoring; Noise; Sleep;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference (BioCAS), 2011 IEEE
  • Conference_Location
    San Diego, CA
  • Print_ISBN
    978-1-4577-1469-6
  • Type

    conf

  • DOI
    10.1109/BioCAS.2011.6107808
  • Filename
    6107808