Title :
Charge-injection waveforms for iridium oxide (AIROF) microelectrodes
Author :
Cogan, S.F. ; Troyk, P.R. ; Ehrlich, J. ; Plante, T.D. ; McCreery, D.B. ; Bullara, L.
Author_Institution :
EIC Labs. Inc., Norwood, MA, USA
Abstract :
The charge-injection limits of activated iridium oxide (AIROF) microelectrodes subjected to charge-balanced biphasic current pulsing are investigated as a function of anodic bias and asymmetry in the cathodic and anodic pulse widths. The use of asymmetric waveforms, in which the charge balancing anodic phase is delivered at a lower current density and longer pulse width, permits the use of anodic biasing to maximize charge-injection capacity. The need for more sophisticated driving waveforms and how these could be implemented in modern ASIC design to achieve optimal charge-injection is discussed.
Keywords :
bioelectric phenomena; biomedical electrodes; charge injection; electrochemistry; iridium compounds; microelectrodes; neurophysiology; IrO2; anodic bias; anodic pulse widths; biphasic current pulsing; cathodic pulse widths; charge-balance; charge-injection waveforms; iridium oxide microelectrodes; neural stimulation; Application specific integrated circuits; Current density; Electrodes; Etching; Insulation; Laboratories; Microelectrodes; Shafts; Space vector pulse width modulation; Wire;
Conference_Titel :
Engineering in Medicine and Biology Society, 2003. Proceedings of the 25th Annual International Conference of the IEEE
Print_ISBN :
0-7803-7789-3
DOI :
10.1109/IEMBS.2003.1279829