• DocumentCode
    2720467
  • Title

    Capabilities and limitations of electromagnetic hyperthermia applicators

  • Author

    Johnson, R.H. ; Preece, A.W. ; Green, J.L.

  • Author_Institution
    Biophys. Group, Bristol Oncology Center, UK
  • fYear
    1995
  • fDate
    34758
  • Firstpage
    42705
  • Lastpage
    42709
  • Abstract
    The heating profiles of different electrically small applicators are shown to be similar and field penetration to be a function of aperture size, frequency and tissue conductivity. Below 100 MHz field penetration of a given applicator is almost independent of low values of conductivity. Single or multiple surface applicators cannot effectively treat deep tumours. It is considered that the practical usefulness of hyperthermia using superficial electromagnetic applicators is limited
  • Keywords
    biomedical equipment; hyperthermia; 100 MHz; aperture size; deep tumours treatment; electrically small applicators; electromagnetic hyperthermia applicators; field penetration; frequency; heating profiles; medical therapeutic technique; multiple surface applicators; single applicators; superficial electromagnetic applicators; tissue conductivity;
  • fLanguage
    English
  • Publisher
    iet
  • Conference_Titel
    Application of Microwaves in Medicine, IEE Colloquium on
  • Conference_Location
    London
  • Type

    conf

  • DOI
    10.1049/ic:19950263
  • Filename
    478218