DocumentCode
2732184
Title
Images reproducibility of an electrical impedance tomography (EIT) prototype. Analysis of the EIT sensibility in rats in pathological in vivo conditions
Author
Gaona, A. ; Aguillon, P. ; Mendoza, Jorge ; Lopez, L. ; Gonzalez, E.
Author_Institution
Laboratorio de Bioingenieria, Inst. Nacional de Psiquiatria, Mexico City, Mexico
Volume
1
fYear
2004
fDate
1-5 Sept. 2004
Firstpage
2296
Lastpage
2298
Abstract
The development of a 16 electrode-electrical impedance tomography (EIT) prototype to be applied in neurological fields such as epilepsy in rats has been previously reported. Approaching residual problems in order to improve its performance, this work reports results about changes made in the system hardware as follows: 1) replacing the current source demultiplexing circuit that could impact on a better spatial localization, and 2) a new current source design that increases the current amplitude up to 5 mArms. System was evaluated by means of: a) image reproducibility starting from 4 test elements in homogeneous conditions; and b) spatial localization evaluation in conductivity perturbation conditions; this feature is evaluated too in preliminary acute in vivo experiments where an epileptic seizure is induced, and an impedance increase is expected. Results show a 95% of proper images for a) analysis. Spatial localization reports improvement up to 20% transversely and 5.5% longitudinally with regard to previous results. In vivo results are lack of interpretation due poor changes obtained in images. In order to conclude or not a reliable correlation between the perturbation measured and the seizure activity, a new definition of grey scale or other changes could be proposed.
Keywords
bioelectric phenomena; biomedical electrodes; demultiplexing; diseases; electric impedance imaging; medical image processing; neurophysiology; conductivity perturbation conditions; current source demultiplexing circuit; current source design; electrical impedance tomography; electrode; epilepsy; image reproducibility; neurology; pathological in vivo conditions; rats; spatial localization; Demultiplexing; Epilepsy; Hardware; Impedance; In vivo; Pathology; Prototypes; Rats; Reproducibility of results; Tomography; Conductivity Perturbation; Current Source; Residual Problems in EIT; Spatial Localization;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
Conference_Location
San Francisco, CA
Print_ISBN
0-7803-8439-3
Type
conf
DOI
10.1109/IEMBS.2004.1403667
Filename
1403667
Link To Document