DocumentCode :
2733167
Title :
A ring-type device for the noninvasive measurement of arterial blood pressure
Author :
Zheng, D.C. ; Zhang, Y.T.
Author_Institution :
Dept. of Electron. Eng., Chinese Univ. of Hong Kong, China
Volume :
4
fYear :
2003
fDate :
17-21 Sept. 2003
Firstpage :
3184
Abstract :
A ring-type device based on photoplethysmogram (PPG) and electrocardiogram (ECG) can be worn by a person at all times for the noninvasive and continuous measurement of arterial blood pressure (BP). A motion-free PPG is very important in order to obtain accurate BP value. In this paper, the optimal finger-ring and circuit designs for stable PPG signal detection and accurate R-wave detection are presented. A series of comparative experiments are conducted on 10 subjects using various ring sensors with different configurations. The experimental results show that the convex-type ring sensor, in which light emitting diode (LED) and infrared (IR) detector are placed at the palm side, and separated by a distance of around 1cm, is the optimal choice for ring-type BP measurement.
Keywords :
biomedical equipment; blood pressure measurement; electrocardiography; infrared detectors; light emitting diodes; medical signal detection; plethysmography; R-wave detection; arterial blood pressure; convex-type ring sensor; electrocardiogram; finger ring-type device; infrared detector; light emitting diode; motion-free PPG signal detection; noninvasive measurement; photoplethysmogram; ring-type BP measurement; Arterial blood pressure; Biomedical measurements; Circuits; Electrocardiography; Infrared detectors; Infrared sensors; Light emitting diodes; Noninvasive treatment; Pressure measurement; Pulse measurements;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2003. Proceedings of the 25th Annual International Conference of the IEEE
ISSN :
1094-687X
Print_ISBN :
0-7803-7789-3
Type :
conf
DOI :
10.1109/IEMBS.2003.1280819
Filename :
1280819