• DocumentCode
    2738862
  • Title

    Determining critical values of calculated parameters within a physiologic model

  • Author

    Hutchinson, G.M.

  • Author_Institution
    GE Healthcare – Information Technologies, Milwaukee, WI, USA
  • Volume
    2
  • fYear
    2004
  • fDate
    1-5 Sept. 2004
  • Firstpage
    3436
  • Lastpage
    3439
  • Abstract
    To determine the critical value of calculated but not readily measurable variables of a physiologic model, we examine clinical studies of similar variables and establish the critical values. The calculated variables considered in this study are the partial pressure of oxygen in the myocardium and in the gray matter of the brain. Partial pressure target values of 29 mmHg for cerebral tissue and 23 mmHg for myocardial tissue were determined to be suitable for a critical threshold. We validated these with a physiologic model test by ensuring that these corresponded to a reasonable value of cardiac output, vital sign commonly used to evaluate patient well being. These values can be used for future analysis of physiologic monitors´ alarm systems when using physiologic models.
  • Keywords
    Physiologic monitors; alarms; physiologic models; Alarm systems; Biomedical monitoring; Blood flow; Brain modeling; Humans; Myocardium; Patient monitoring; Performance evaluation; Pressure measurement; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
  • Print_ISBN
    0-7803-8439-3
  • Type

    conf

  • DOI
    10.1109/IEMBS.2004.1403964
  • Filename
    1403964