Title :
End tidal carbon dioxide measurement using an electro acoustic sensor
Author :
Folke, M. ; Hök, B. ; Ekström, M. ; Bäcklund, Y.
Author_Institution :
Department of Electrical Engineering, Mälardalen University, Västerås, Sweden
Volume :
2
fYear :
2004
fDate :
1-5 Sept. 2004
Firstpage :
3452
Lastpage :
3455
Abstract :
End tidal carbon dioxide measurement with an electro-acoustic sensor is demonstrated. The sensor consists of an acoustic resonator coupled to a low cost electro-acoustic element. By simultaneous measurements with a reference sensor, the new device was tested on subjects performing exercise, hypo- and hyperventilation whereby the CO2concentration ranged from 2.1 to 7.0 kPa. The output from the experimental device correlated well with the reference CO2readings with a correlation coefficient of 0.976. Response time for expiration less than 0.8 seconds was noted. The new device could be useful in situations where selectivity to other gases is not important.
Keywords :
End tidal; capnometry; carbon dioxide; electro acoustic sensor; Acoustic devices; Acoustic measurements; Acoustic sensors; Acoustic testing; Carbon dioxide; Chemical elements; Costs; Delay; Gas detectors; Performance evaluation;