DocumentCode :
2743754
Title :
Investigation of systematic errors of an AC Josephson Voltage Standard
Author :
Filipski, Piotr S.
Author_Institution :
Nat. Res. Council Canada - INMS, Ottawa, ON, Canada
fYear :
2012
fDate :
1-6 July 2012
Firstpage :
36
Lastpage :
37
Abstract :
This paper presents results of experiments quantifying two systematic errors associated with the calibrations by use of an AC Josephson Voltage Standard: the change in the voltage leads error with the change in the level of probe immersion in the liquid helium, and the effect of the internal inductance of a Josephson array. The measurements were performed at frequencies 100 kHz and above, however, they can be used in evaluating an error budget at lower frequencies.
Keywords :
Josephson effect; calibration; inductance; liquid helium; measurement errors; measurement standards; superconducting arrays; voltage measurement; AC Josephson voltage standard; Josephson array; calibration; frequency 100 kHz; inductance effect; liquid helium; probe immersion; systematic error; voltage lead error; Arrays; Helium; Inductance; Liquids; Probes; Standards; Voltage measurement; Digital-analog conversion; Josephson arrays; measurement uncertainty; quantization; signal synthesis; standards; voltage measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location :
Washington, DC
ISSN :
0589-1485
Print_ISBN :
978-1-4673-0439-9
Type :
conf
DOI :
10.1109/CPEM.2012.6250646
Filename :
6250646
Link To Document :