DocumentCode
2743903
Title
A bias source for the voltage reference of the BIPM watt balance
Author
Solve, S. ; Chayramy, R. ; Picard, A. ; Kiss, A. ; Fang, H. ; Robertsson, L. ; De Mirandés, E. ; Stock, M.
Author_Institution
Bur. Int. des Poids et Mesures (BIPM), Sevres, France
fYear
2012
fDate
1-6 July 2012
Firstpage
52
Lastpage
53
Abstract
We have developed a programmable current bias source to set the voltage output of a Josephson Voltage Standard that will be operated on the watt balance experiment. The device has been designed to bias separately all of the 13 independent segments of the Josephson array in order to be able to track the induced voltage across the moving coil during the constant velocity and acceleration phases. It is powered from Lithium-ion batteries in order to avoid electrical offsets originating from ground loops and to allow long series of measurements.
Keywords
coils; measurement standards; reference circuits; secondary cells; voltage measurement; BIPM watt balance experiment; Josephson voltage standard array; acceleration phase; constant velocity; electrical offset; ground loop; induced voltage tracking; lithium-ion battery; moving coil; programmable current bias source; voltage reference; Arrays; Batteries; Coils; Current measurement; Standards; Time measurement; Voltage measurement; Josephson Voltage Standard; Programmable Bias Current source; watt balance experiment;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location
Washington, DC
ISSN
0589-1485
Print_ISBN
978-1-4673-0439-9
Type
conf
DOI
10.1109/CPEM.2012.6250656
Filename
6250656
Link To Document