• DocumentCode
    2743903
  • Title

    A bias source for the voltage reference of the BIPM watt balance

  • Author

    Solve, S. ; Chayramy, R. ; Picard, A. ; Kiss, A. ; Fang, H. ; Robertsson, L. ; De Mirandés, E. ; Stock, M.

  • Author_Institution
    Bur. Int. des Poids et Mesures (BIPM), Sevres, France
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    52
  • Lastpage
    53
  • Abstract
    We have developed a programmable current bias source to set the voltage output of a Josephson Voltage Standard that will be operated on the watt balance experiment. The device has been designed to bias separately all of the 13 independent segments of the Josephson array in order to be able to track the induced voltage across the moving coil during the constant velocity and acceleration phases. It is powered from Lithium-ion batteries in order to avoid electrical offsets originating from ground loops and to allow long series of measurements.
  • Keywords
    coils; measurement standards; reference circuits; secondary cells; voltage measurement; BIPM watt balance experiment; Josephson voltage standard array; acceleration phase; constant velocity; electrical offset; ground loop; induced voltage tracking; lithium-ion battery; moving coil; programmable current bias source; voltage reference; Arrays; Batteries; Coils; Current measurement; Standards; Time measurement; Voltage measurement; Josephson Voltage Standard; Programmable Bias Current source; watt balance experiment;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6250656
  • Filename
    6250656