DocumentCode
2744237
Title
Waveguide 50 – 110 GHz power sensor calibration system by direct comparison transfer
Author
Shan, Yueyan ; Meng, Yu Song ; Neo, Hoon
Author_Institution
Nat. Metrol. Centre (NMC), A*STAR (Agency for Sci., Singapore, Singapore
fYear
2012
fDate
1-6 July 2012
Firstpage
94
Lastpage
95
Abstract
This paper proposes a different modeling technique for analyzing the microwave power sensor calibration by direct comparison transfer. The model is established by using signal flow graph together with non-touching loop rule analyses. The realization of the banded calibration system from 50 GHz to 110 GHz by WR15 and WR10 connections is also described.
Keywords
calibration; coaxial waveguides; flow graphs; microwave detectors; millimetre wave detectors; sensors; WR10 connections; WR15 connections; coaxial waveguide power sensor calibration system; direct comparison transfer; frequency 50 GHz to 110 GHz; microwave power sensor calibration; nontouching loop rule analyses; signal flow graph; Analytical models; Calibration; Equations; Flow graphs; Mathematical model; Microwave theory and techniques; Power measurement; Direct comparison transfer method; measurement modeling; measurement uncertainty; microwave power sensor calibration; waveguide connection;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location
Washington, DC
ISSN
0589-1485
Print_ISBN
978-1-4673-0439-9
Type
conf
DOI
10.1109/CPEM.2012.6250677
Filename
6250677
Link To Document