Title :
PVDF hydrophone with liquid electrodes for shock wave measurements [in lithotripsy]
Author_Institution :
INSERM, Lyon, France
Abstract :
A novel hydrophone with nonmetallic electrodes is proposed. The polyvinylidene fluoride (PVDF) film hydrophone is polarized over a small area at its center. The electrical charges on this sensitive area are then driven electrically to the external leads through an electrolyte. With a thin 9-mm PVDF foil and an unmetallized sensitive area of 1 mm, the resulting hydrophone has a sensitivity of 12 mV/MPa over a bandwidth of about 20 MHz. More than 105 focused shock pulses of 30 MPa have been measured without any significant decrease in hydrophone sensitivity
Keywords :
acoustic intensity measurement; biomedical equipment; biomedical ultrasonics; hydrophones; kidney; polymer films; radiation therapy; shock wave effects; ultrasonic measurement; ultrasonic transducers; 20 MHz; PVDF hydrophone; liquid electrodes; lithotripters; sensitivity; shock wave measurements; Acoustic measurements; Biomembranes; Electric shock; Electrodes; Lithotripsy; Lithotriptors; Pressure measurement; Shock waves; Sonar equipment; Ultrasonic variables measurement;
Conference_Titel :
Ultrasonics Symposium, 1990. Proceedings., IEEE 1990
Conference_Location :
Honolulu, HI
DOI :