• DocumentCode
    2747186
  • Title

    A bilateral comparison between NIST quantum-based power standard and NRC current-comparator-based power standard

  • Author

    Waltrip, B.C. ; Nelson, T.L. ; So, E. ; Angelo, D.

  • Author_Institution
    Phys. Meas. Lab., Quantum Meas. Div., Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    203
  • Lastpage
    204
  • Abstract
    The results of a bilateral comparison on active/reactive power meter calibrations between the National Institute of Standards and Technology (NIST) and the National Research Council (NRC) will be presented. The comparison was implemented using a transfer standard consisting of a highly stable commercial sampling-type power/energy meter. The measurements were made at 120 V, 5A, 50 Hz and 60 Hz, at power factors of 1.0, 0.5 lead and lag, and 0.0 lead and lag.
  • Keywords
    calibration; power meters; reactive power; NIST quantum-based power standard; NRC current-comparator-based power standard; National Research Council; active-reactive power meter calibrations; bilateral comparison; current 5 A; frequency 50 Hz; frequency 60 Hz; high stable commercial sampling-type power-energy meter; voltage 120 V; Bridge circuits; Current measurement; Laboratories; NIST; Reactive power; Uncertainty; Current-comparator; power bridge; power standard; quantum; uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6250872
  • Filename