DocumentCode
2748970
Title
Cross capacitor with movable main electrodes
Author
Semenov, Y.P. ; Shvedov, O.A.
Author_Institution
D.I. Mendeleyev Inst. for Metrol. (VNIIM), St. Petersburg, Russia
fYear
2012
fDate
1-6 July 2012
Firstpage
396
Lastpage
397
Abstract
The method is proposed for changing the effective length the electrodes of calculable cross capacitor (CC) by use of sectioned detector electrodes and their movement along the axis of capacitor electrodes. This method allows to eliminate a number of shortcomings of conventional CCs, in which the length is changed by moving the guard electrodes. Guard electrodes disturb the field that is described in Thompson-Lampard theorem and multiply the effect of local imperfections on the electrodes and misalignment of the electrodes.
Keywords
capacitance measurement; capacitors; electrodes; capacitor electrode; cross capacitor; effective electrode length; guard electrode; movable main electrodes; sectioned detector electrode; Accuracy; Capacitance; Capacitance measurement; Capacitors; Detectors; Distortion measurement; Electrodes; Cross capacitor; end effect; guard electrode; interferometer; misalignment;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location
Washington, DC
ISSN
0589-1485
Print_ISBN
978-1-4673-0439-9
Type
conf
DOI
10.1109/CPEM.2012.6250970
Filename
6250970
Link To Document