• DocumentCode
    2749838
  • Title

    A simple build-up method for the DC voltage scale of a source

  • Author

    Early, Murray ; Christian, Laurie ; Andersson, Bengt-Olof ; Gunnarsson, Ove ; Rydler, Karl-Erik

  • Author_Institution
    Meas. Stand. Lab. of New Zealand (MSL), Lower Hutt, New Zealand
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    498
  • Lastpage
    499
  • Abstract
    We report on the performance of a simple method, known as the Reference Step Method, to build up the dc voltage scale of a source. The key advantages of this method are that it relies on equipment that is readily available in most high-level laboratories and can be fully automated. The Type A uncertainties achieved by this approach can be less than 0.1 μV/V. This performance is achieved because of the excellent differential linearity of the DMM used in the method. We are validating the accuracy of the method by directly comparing this approach with more traditional resistive divider techniques.
  • Keywords
    calibration; voltage measurement; DC voltage scale calibration; DMM differential linearity; build-up method; high-level laboratories; reference step Method; resistive divider techniques; type A uncertainties; Accuracy; Calibration; Gain measurement; Instruments; Linearity; Measurement standards; Voltage measurement; DC voltage ratio; dc build-up; dc range calibration; dc voltage scale;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6251021
  • Filename
    6251021