DocumentCode :
2750747
Title :
Automated capacitance bridge for calibration of capacitors with nominal value from 10 nF up to 10 mF
Author :
Kucera, Jan ; Funck, Torsten ; Melcher, Jürgen
Author_Institution :
Phys.-Tech. Bundesanstalt, Braunschweig, Germany
fYear :
2012
fDate :
1-6 July 2012
Firstpage :
596
Lastpage :
597
Abstract :
This paper describes an automatized capacitance bridge (KRMS-C) developed for calibration of dielectric loss capacitors in the wide range from 10 nF up to 10 mF at frequencies from 50 Hz up to 10 kHz. The system is based on technique, where voltage drops across known and unknown impedance are measured and when measurement conditions are fulfilled, the value of the unknown impedance can be evaluated. The arrangement of the KRMS-C allows for performing calibration of 4-TP capacitors and with appropriate adaptors also 3-T and 5-T capacitors. Due to the integrated scanner/multiplexer, the calibration of up to four capacitors is possible in a single setup.
Keywords :
calibration; capacitors; dielectric losses; 3-T capacitors; 4-TP capacitors; 5-T capacitors; KRMS-C; adaptors; automated capacitance bridge; calibration; dielectric loss capacitors; integrated scanner-multiplexer; measurement conditions; voltage drops; Bridge circuits; Calibration; Capacitance; Capacitors; Resistors; Uncertainty; Voltage measurement; capacitance; measurement techniques; sampling system;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location :
Washington, DC
ISSN :
0589-1485
Print_ISBN :
978-1-4673-0439-9
Type :
conf
DOI :
10.1109/CPEM.2012.6251070