• DocumentCode
    2750747
  • Title

    Automated capacitance bridge for calibration of capacitors with nominal value from 10 nF up to 10 mF

  • Author

    Kucera, Jan ; Funck, Torsten ; Melcher, Jürgen

  • Author_Institution
    Phys.-Tech. Bundesanstalt, Braunschweig, Germany
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    596
  • Lastpage
    597
  • Abstract
    This paper describes an automatized capacitance bridge (KRMS-C) developed for calibration of dielectric loss capacitors in the wide range from 10 nF up to 10 mF at frequencies from 50 Hz up to 10 kHz. The system is based on technique, where voltage drops across known and unknown impedance are measured and when measurement conditions are fulfilled, the value of the unknown impedance can be evaluated. The arrangement of the KRMS-C allows for performing calibration of 4-TP capacitors and with appropriate adaptors also 3-T and 5-T capacitors. Due to the integrated scanner/multiplexer, the calibration of up to four capacitors is possible in a single setup.
  • Keywords
    calibration; capacitors; dielectric losses; 3-T capacitors; 4-TP capacitors; 5-T capacitors; KRMS-C; adaptors; automated capacitance bridge; calibration; dielectric loss capacitors; integrated scanner-multiplexer; measurement conditions; voltage drops; Bridge circuits; Calibration; Capacitance; Capacitors; Resistors; Uncertainty; Voltage measurement; capacitance; measurement techniques; sampling system;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6251070
  • Filename
    6251070