Title :
Development of 171Yb optical lattice clock at KRISS
Author :
Park, Chang Yong ; Yu, Dai-Hyuk ; Lee, Won-Kyu ; Park, Sang Eon ; Kwon, Taeg Yong ; Lee, Sang-Bum ; Mun, Jongchul
Author_Institution :
Center for Time & Freq., Korea Res. Inst. of Stand. & Sci., Daejeon, South Korea
Abstract :
We developed an 171Yb optical lattice clock and measured the absolute frequency of the optical clock transition 1S0 (F = 1/2) - 3P0 (F = 1/2). The frequency was determined at 518 295 836 590 865.7 (9.2). The uncertainty is equivalent to 1.8×10-14 relatively. The magic wavelength of lattice laser was founded at 394 798.18(79) Hz.
Keywords :
atomic clocks; frequency measurement; optical lattices; ytterbium; KRISS; Yb; absolute frequency measurement; frequency 394798.18 Hz; lattice laser magic wavelength; optical lattice clock transition; Atom optics; Atomic clocks; Atomic measurements; Frequency measurement; Laser transitions; Lattices; Optical clock; clock transition of 171Yb atoms; magic wavelength; optical lattice clock; precision measurements;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location :
Washington, DC
Print_ISBN :
978-1-4673-0439-9
DOI :
10.1109/CPEM.2012.6251080