• DocumentCode
    2751820
  • Title

    First experimental results of the frequency dependence of a Vacuum-Gap Capacitor between 1 kHz and 0.02 Hz

  • Author

    Rietveld, Gert ; Zimmerman, Neil M. ; Van den Brom, Belko E.

  • Author_Institution
    VSL, Delft, Netherlands
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    708
  • Lastpage
    709
  • Abstract
    This paper describes the first results of an experiment aiming to determine the frequency dependence of a cryogenic vacuum-gap capacitor between 0.01 Hz (`DC´) and 1 kHz (AC). The AC-`DC´ difference of the vacuum-gap capacitor is measured as a function of temperature, and the measured change of around 23 μF/F between 273 K and 4 K agrees well with an extrapolation of earlier measurements on the same capacitor between 100 Hz and 1 kHz. The results also seem to support estimates that the AC-`DC´ difference of this capacitor is well below 1 μF/F at 4 K.
  • Keywords
    capacitance measurement; capacitors; cryogenics; AC-DC difference; cryogenic vacuum-gap capacitor; frequency 1 kHz to 0.01 Hz; temperature 273 K to 4 K; Extrapolation; Fires; calibration; capacitance measurement; capacitors; charge measurement; cryogenic vacuum-gap capacitor; current measurement; electrometer; frequency dependence; quantum current standard;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6251126
  • Filename