A new design and evaluation technique of a microcalorimeter
Author :
Cui, Xiaohai ; Li, Yong ; Gao, Xiaoxun
Author_Institution :
Nat. Inst. of Metrol. of China, Beijing, China
fYear :
2012
fDate :
1-6 July 2012
Firstpage :
736
Lastpage :
738
Abstract :
A new design and evaluation technique of a rectangular waveguide microcalorimeter is presented. Its design is based on the NIST new WR-15 microcalorimeter design[1], but an important modification was made so that the principle of the measurement was changed. In this design, not only can take the original evaluation technique, but also take the new method to obtain the final data-reference standard´s effective efficiency. By the new evaluation technique, we don´t need measure the S-parameter of the trans-line in the microcalorimeter which is the main uncertainty source in obtaining the thermistor mount effective efficiency.
Keywords :
S-parameters; calorimeters; measurement standards; measurement uncertainty; rectangular waveguides; thermistors; transmission lines; S-parameter; WR-15 microcalorimeter design; data reference standard; measurement uncertainty; rectangular waveguide microcalorimeter; thermistor; transmission line; Equations; NIST; Power measurement; Radio frequency; Temperature measurement; Thermistors; Microcalorimeter; microwave power; millimeter-wave power; primary standards;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location :
Washington, DC
ISSN :
0589-1485
Print_ISBN :
978-1-4673-0439-9
Type :
conf
DOI :
10.1109/CPEM.2012.6251140
Filename :
6251140
Link To Document :
بازگشت