Abstract :
Shock wave generators are now currently used for the treatment of renal stones. In all these generators the focal zone is determined by their geometrical parameters. We propose a piezo-composite shock wave generator with electronic focusing. The system is composed of a bidimensional array and its electronic hardware. The array is composed of 274 independent piezo-composite transducers arranged in a spherical shell 28 cm in diameter and focused at 190 mm. The electronic hardware includes 274×6 kV-impulse generators. The inter-delay of each channel can be adjusted between 100 ns to 100 μs by steps of 100 ns. Electronic focusing is efficient in an ellipsoidal region of about 4 cm in diameter and 12 cm in length, this volume is large enough to follow the displacement of the renal stones due to breathing during the treatment. In vitro results show that it is possible to destroy plaster balls located everywhere in this region
Keywords :
biomedical ultrasonics; kidney; piezoelectric transducers; pressure measurement; radiation therapy; shock waves; ultrasonic focusing; ultrasonic transducer arrays; 100 ns to 100 mus; 12 cm; 190 mm; 28 cm; 30 cm; 30 cm diameter bidimensional array; 4 cm; 6 kV; breathing; channel inter-delay; electronic focusing; electronic hardware; elementary transducers; ellipsoidal region; extracorporeal shock wave lithotripsy; focal zone; geometrical parameters; impulse generators; in vitro results; piezo-composite shock wave generator; piezo-composite transducers; piezoelectric array; plaster balls; renal stone displacement; renal stone treatment; shock wave beam steering; shock wave generators; spherical shell; Acoustic arrays; Acoustic beam steering; Biomedical applications of acoustic radiation; Kidneys; Piezoelectric transducers; Pressure measurement; Shock waves;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 1994. Proceedings., 1994 IEEE
Conference_Location :
Cannes, France
Print_ISBN :
0-7803-2012-3
Type :
conf
DOI :
10.1109/ULTSYM.1994.401940
Filename :
401940
Link To Document :
بازگشت