• DocumentCode
    2770013
  • Title

    Trace explosives detection by micro differential thermal analysis

  • Author

    Olsen, Jesper ; Senesac, Larry ; Thundat, Thomas ; Boisen, Anja

  • Author_Institution
    Tech. Univ. of Denmark, Lyngby, Denmark
  • fYear
    2011
  • fDate
    23-27 Jan. 2011
  • Firstpage
    984
  • Lastpage
    987
  • Abstract
    In this abstract a new method for detecting trace explosive compounds using a micro differential total analysis system (DTA) is presented. Trace explosive detection has become a predominant research activity in recent years and there is a demand for fast reliable screening methods to secure public safety. Micro DTA facilitates recording of a calorimetric profile for nanograms of materials. The profile comprises a combination of sublimation, melting, evaporation and deflagration for the sample material. Using this profile it is possible to distinguish explosives from non explosives and to have selectivity between explosives.
  • Keywords
    chemical sensors; differential thermal analysis; explosives; micro differential thermal analysis; micro differential total analysis system; trace explosives detection; Biomembranes; Explosives; Heating; Resistance; Resistors; Temperature measurement; Temperature sensors;