Title :
Electromagnetic interference coupling into cardiac pacemaker electrodes
Author :
Schenke, Stefan ; Sabath, Frank ; Sutter, Florian ; Clemens, Markus ; Dickmann, Stefan
Author_Institution :
Fak. fur Elektrotech., Helmut-Schmidt-Univ., Hamburg
fYear :
2008
fDate :
8-12 Sept. 2008
Firstpage :
1
Lastpage :
5
Abstract :
Cardiac pacemakers (PMKs) are vital for many persons sufferering from heart diseases, and a malfunction of those devices can have terrible consequences for their wearers. The main source of malfunctions is the PMK´s electrode which acts linke an antenna. An incident electromagnetic field can couple into the electrode and cause an interference voltage at the PMK´s input connector. This signal can be misinterpreted as the electrocardiogram (ECG) of the heart and mislead the PMK to wrong reactions. This paper presents the results of series simulations for calculating the maximum input voltage as a function of the electical field strength of a plane wave penetrating the human body.
Keywords :
electrocardiography; electromagnetic fields; electromagnetic interference; pacemakers; ECG; antenna; cardiac pacemaker electrodes; electrocardiogram; electromagnetic interference coupling; heart diseases; human body; incident electromagnetic field; Cardiac disease; Connectors; Electrocardiography; Electrodes; Electromagnetic coupling; Electromagnetic fields; Electromagnetic interference; Heart; Pacemakers; Voltage; cardiac pacemaker; disturbance; electrode; interference voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electromagnetic Compatibility - EMC Europe, 2008 International Symposium on
Conference_Location :
Hamburg
Print_ISBN :
978-1-4244-2737-6
Electronic_ISBN :
978-1-4244-2737-6
Type :
conf
DOI :
10.1109/EMCEUROPE.2008.4786824
Filename :
4786824
Link To Document :
بازگشت