DocumentCode :
2775675
Title :
Intraoperative Evaluation of the Spiral Nerve Cuff Electrode for a Standing Neuroprosthesis
Author :
Polasek, Katharine H. ; Schiefer, Matthew A. ; Pinault, Gilles C. ; Triolo, Ronald J. ; Tyler, Dustin J.
Author_Institution :
Dept. of Biomed. Eng., Case Western Reserve Univ., Cleveland, OH
fYear :
2007
fDate :
2-5 May 2007
Firstpage :
89
Lastpage :
92
Abstract :
Evaluation of the spiral nerve cuff electrode on the proximal femoral nerve for a standing neuroprosthesis was performed intraoperatively in 4 subjects. The mean stimulation threshold was 17.7 plusmn 12 nC, similar to stimulation thresholds in upper extremity nerves. The femoral nerve was found to have an oblong cross section with an average width of 9 mm and height of 1-2 mm. In all 4 subjects where data was recorded, selective activation of at least one hip flexor (rectus femoris or sartorius) was possible. For a standing neuroprosthesis, knee extension without hip flexion is desired. A more selective electrode is needed for this application.
Keywords :
biological organs; biomedical electrodes; neural nets; neuromuscular stimulation; prosthetics; functional electrical stimulation; hip flexion; hip flexor; implantable electrodes; intraoperative evaluation; knee extension; mean stimulation threshold; nerve electrodes; neuroprosthesis; peripheral nerve; proximal femoral nerve; spinal cord injury; spiral nerve cuff electrode; Biomedical engineering; Electrodes; Extremities; Hip; Knee; Muscles; Neural engineering; Neuromuscular stimulation; Orthopedic surgery; Spirals; Functional Electrical Stimulation (FES); implantable electrodes; nerve electrodes; peripheral nerve; spinal cord injury;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Neural Engineering, 2007. CNE '07. 3rd International IEEE/EMBS Conference on
Conference_Location :
Kohala Coast, HI
Print_ISBN :
1-4244-0792-3
Electronic_ISBN :
1-4244-0792-3
Type :
conf
DOI :
10.1109/CNE.2007.369619
Filename :
4227224
Link To Document :
بازگشت