Abstract
Power-line interference is a major problem in High Resolution Biopotential Measurements, mostly because of capacitive coupling. This paper describes the different coupling paths and the parameters involved in interference reduction. One of the parameters that determine the amplitude of low frequency interference in biopotential recordings is skin-electrode impedance but reducing it, particularly in multichannel measurements is not easy. The authors propose an active electrode (based on commercial electrodes) able to reduce power-line interference by about 20 dB
Keywords
bioelectric potentials; biomedical equipment; biomedical measurement; electrodes; interference (signal); interference suppression; 20 dB; capacitive coupling; commercial electrodes; coupling paths; high resolution biopotential measurements; low frequency interference; medical instrumentation; multichannel measurements; power line interference reduction; simple active electrode; skin-electrode impedance; Biomedical electrodes; Capacitance; Electrocardiography; Impedance; Interference; Joining processes; Operational amplifiers; Power system modeling; Voltage; Wires;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1996. Bridging Disciplines for Biomedicine. Proceedings of the 18th Annual International Conference of the IEEE
Conference_Location
Amsterdam
Print_ISBN
0-7803-3811-1
Type
conf
DOI
10.1109/IEMBS.1996.656864
Filename
656864
Link To Document