DocumentCode :
2780653
Title :
C-shaped embedded polymer waveguide for evanescent field absorption based lab on a chip biosensor
Author :
Prabhakar, Amit ; Mukherji, Soumyo
Author_Institution :
Dept. of Biosci. & Bioeng., IIT Bombay, Mumbai, India
fYear :
2010
fDate :
16-18 Dec. 2010
Firstpage :
67
Lastpage :
70
Abstract :
In this report a microfluidic chip with embedded C-shaped waveguides was fabricated and study was performed to determine its biosensing capabilities. The general fabrication procedure to couple the microchannel network with C-bend waveguides was a single step process in which the photo resist SU-8 layer was patterned into optical structures (C-bend waveguides), fiber-to-waveguide coupler and microchannel network structures (Fig. 2) using single mask (Fig. 1). The device was tested for sensitivity to refractive index variation in the analyte (by passing sucrose solutions of varying concentrations) as well as evanescent wave absorption (by passing various concentrations of Methylene Blue dye solutions) as explained in Fig 3 and Fig 4. These embedded C-shaped waveguide probes demonstrated biosensing with the help of Human IgG (HIgG) - FITC tagged goat anti IgG(GaHIgG) biomolecules as bioreceptor-analyte pair. For HIgG immobilization on C-shaped SU-8 waveguide the glycine was used as crosslinkers, before applying the carbodiimide/succinimide chemistry over the waveguide surface. The analyte (FITC tagged GaHIgG) was passed through the micro-channel embedded with the C-shaped waveguide. The antigen- antibody interaction and binding results in to the absorption of evanescent field by FITC tagged GaHIgG bio-molecule, causing a drop in light intensity output of the waveguide which was detected and calibrated. The results indicated that C-shaped waveguides can be used easily for more sensitive bio sensing of biomolecules either by evanescent wave absorption or by changes associated with refractive index in the microenvironment around a waveguide.
Keywords :
biosensors; lab-on-a-chip; microfluidics; photoresists; polymers; waveguides; C-shaped embedded polymer waveguide; FITC tagged GaHIgG; HIgG immobilization; Human IgG; Methylene Blue dye; bioreceptor-analyte pair; evanescent field absorption; glycine; lab-on-a-chip biosensor; microchannel network; photo resist SU-8 layer; Biomedical optical imaging; Optical fiber networks; Optical fiber sensors; Optical fibers; C-shaped polymer waveguide; FITC tagged goat anti IgG(GaHIgG); Human Immunoglobin G (HIgG); biosensor; evanescent wave absorption; microchannel; refractive index sensitivity;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Systems in Medicine and Biology (ICSMB), 2010 International Conference on
Conference_Location :
Kharagpur
Print_ISBN :
978-1-61284-039-0
Type :
conf
DOI :
10.1109/ICSMB.2010.5735347
Filename :
5735347
Link To Document :
بازگشت