Title :
Eddy current compensation using screened gradients
Author :
Rawson, M.H. ; Judson, R.A. ; Ellis, R.E. ; Pittard, S. ; Vennart, W.
Author_Institution :
Dept. of Phys., Exeter Univ., UK
Abstract :
In magnetic resonance imaging using superconducting magnet eddy currents are induced in the magnet dewars during the switching of magnetic field gradients. This leads to a loss of amplitude and linearity in the resulting gradients. Three methods of compensation are available, namely to work with the gradient set as far as possible from the magnet dewars, to use software or hardware to apply pre-emphasis and finally to use active screened gradient sets. None of these techniques give perfect compensation, but clearly combinations can be used to obtain optimum performance
Keywords :
biomedical NMR; biomedical equipment; coils; compensation; eddy currents; superconducting magnets; active screened gradient sets; eddy current compensation; magnetic resonance imaging; medical diagnostic imaging; optimum performance; superconducting magnet;
Conference_Titel :
Technical Developments Relating to Clinical NMR in the UK, IEE Colloquium on
Conference_Location :
London