DocumentCode
278483
Title
A novel method to determine and control power delivered during laser angioplasty
Author
Smith, R.E. ; Somkuarnpanit, S. ; Milnes, P. ; Edwards, D. ; Mitchell, D.C. ; Wood, R.F.M.
Author_Institution
St. Bartholomew´´s Hospital, London, UK
fYear
1991
fDate
33403
Firstpage
42461
Lastpage
42464
Abstract
For accurate measurements the in-line powermeter (ILP) described requires standardising against a bolometer each and every time it is used, since the registered power varies with each fibre placement. However, fibre alignment may be accomplished using the ILP alone. The combination of the laser output powermeter (LOP) and the ILP provides accurate and inexpensive control of laser power during angioplasty. Both dye degradation and damage to the proximal end of delivery devices may be identified. The power delivered can thus be accurately controlled without removing the delivery device from the patient, thereby improving safety and efficiency
Keywords
biomedical equipment; fibre optics; laser applications in medicine; power control; radiation therapy; surgery; dye degradation; fibre alignment; in-line powermeter; laser angioplasty; laser output powermeter; laser power control;