Title : 
Trap detectors and their properties
         
        
        
            Author_Institution : 
Div. of Quantum Metrol., NPL, Teddington, UK
         
        
        
        
        
        
            Abstract : 
The author describes a new solid state detector which fulfils all the requirements of the ideal detector for optical radiation measurements. This `trap´ detector is shown to be spatially uniform, linear, stable and with a predictable relative spectral response so that it can be used to realise a relative spectral response scale with an uncertainty of less than 0.05%. The detector is also shown to be capable of disseminating an absolute spectral responsivity scale with an uncertainty of less than 0.02% at specific laser wavelengths and independently realising such a scale with an uncertainty of better than 0.05%
         
        
            Keywords : 
measurement standards; photodetectors; photodiodes; ideal detector; optical radiation measurements; photodiodes; radiometric transfer standard; relative spectral response scale; solid state detector; spatially uniform; standard detector; trap detectors; visible spectral range;
         
        
        
        
            Conference_Titel : 
Developments in near Infra-Red Visible Ultraviolet Detectors, IEE Colloquium on
         
        
            Conference_Location : 
London