Abstract :
It is shown that traceable Q measurements on a stable inductor can provide the basis for dielectric measurements in the frequency range 1-30 MHz. It is also possible to adapt computer controlled four terminal impedance measuring techniques by coupling them with a resonance adaptor and suitable test fixture, so that dielectric measurements may be made to ±2% in permittivity and ±5μ rad in dielectric loss
Keywords :
Q-factor measurement; computerised instrumentation; dielectric loss measurement; inductors; permittivity measurement; 1 to 30 MHz; computer controlled; dielectric loss; dielectric measurements; four terminal impedance measuring techniques; permittivity; resonance adaptor; resonance methods; stable inductor; test fixture; traceable Q measurements;