Title : 
Effect of gap length on effective field strength
         
        
            Author : 
Kojima, A. ; Sato, S. ; Iijima, N. ; Sone, M. ; Mitsui, H.
         
        
            Author_Institution : 
Dept. of Electr. & Electron. Eng., Musashi Inst. of Technol., Tokyo, Japan
         
        
        
        
        
            Abstract : 
A micro-channel plate (MCP) was used to measure the minute emission current within the range of 1×10-19 A to 1×10-14 A in a high vacuum (1.3×10-5 Pa). By a Fowler-Nordheim plot, it is found that the effective field strength for initial electron emission is mostly constant between 107 and 108 V/cm under any gap length. The effective field strength is held constant between 107 and 10 8 V/cm, independent of the current under any gap length. With this result, a model which shows the choice of minute projection is suggested. In order to certify this model, an MCP phosphor-anode is used. As a result, it is found that the minute projections which emit electrons melt with heat and are chosen when the voltage rises
         
        
            Keywords : 
electric current measurement; electron field emission; image intensifiers; 1.3E-5 Pa; 1E-19 to 1E-14 A; Fowler-Nordheim plot; MCP phosphor-anode; effective field strength; emission current; gap length; initial electron emission; micro-channel plate; minute projections; Assembly; Brightness; Cathodes; Current measurement; Electric variables measurement; Electrodes; Electron emission; Iron; Vacuum technology; Voltage;
         
        
        
        
            Conference_Titel : 
Properties and Applications of Dielectric Materials, 1991., Proceedings of the 3rd International Conference on
         
        
            Conference_Location : 
Tokyo
         
        
            Print_ISBN : 
0-87942-568-7
         
        
        
            DOI : 
10.1109/ICPADM.1991.172194