DocumentCode :
279696
Title :
Considerations for the calibration of a broadband, isotropic, E-field sensing system
Author :
Matloubi, K.
Author_Institution :
Electro-Mech. Co., Austin, TX, USA
fYear :
1990
fDate :
32897
Firstpage :
42370
Lastpage :
42373
Abstract :
Describes a method of calibration of a broadband, isotropic, E-field sensing system. A system level design and calibration method is utilized. The calibration dynamic and frequency ranges are 2 to 500 V/m and 1 MHz to 18 GHz, respectively. Calibration is performed in two separate environments; in a transverse electromagnetic (TEM) cell in the frequency range of 1 MHz to 1 GHz and in an anechoic chamber in the range of 1 GHz to 18 GHz
Keywords :
calibration; electric field measurement; 1 MHz to 18 GHz; 2 to 500 V/m; E-field sensing system; TEM; anechoic chamber; broadband; calibration; cell; electric field measurement; isotropic; transverse electromagnetic;
fLanguage :
English