DocumentCode :
2802322
Title :
A concept of cylindrical Compton camera for SPECT
Author :
Bolozdynya, A. ; Ordonez, C.E. ; Chang, W.
Author_Institution :
Rush-Presbyterian-St. Luke´´s Med. Center, Chicago, IL, USA
Volume :
2
fYear :
1997
fDate :
9-15 Nov 1997
Firstpage :
1047
Abstract :
A novel concept of a Compton camera for SPECT is proposed. The camera consists of two concentric cylindrical detectors that operate in coincidence. The camera is triggered by Compton scatter events in the inner (scatter) detector. The scattered photons are then detected in the outer (absorption) detector. If both detectors have excellent energy and three-dimensional (3D) spatial resolutions, such a camera could make 3D imaging possible with appropriate reconstruction algorithms. In this paper, the authors estimate the potential performance of the cylindrical Compton camera that is made up of argon-gas or silicon-based scatter detector and a xenon-gas absorption detector
Keywords :
Compton effect; biomedical equipment; cameras; single photon emission computed tomography; Ar; Si; Xe; absorption detector; argon gas-based scatter detector; cylindrical Compton camera concept; energy resolution; medical diagnostic imaging; medical instrumentation; nuclear medicine; reconstruction algorithms; scatter detector; silicon-based scatter detector; three-dimensional spatial resolution; xenon-gas absorption detector; Absorption; Cameras; Collimators; Detectors; Electromagnetic scattering; Event detection; Optical imaging; Particle scattering; Single photon emission computed tomography; Spatial resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1997. IEEE
Conference_Location :
Albuquerque, NM
ISSN :
1082-3654
Print_ISBN :
0-7803-4258-5
Type :
conf
DOI :
10.1109/NSSMIC.1997.670489
Filename :
670489
Link To Document :
بازگشت