Title :
Large scale phased array and system for treatment of deep seated tissue
Author :
Daum, D.R. ; Hynynen, K.
Author_Institution :
Brigham & Women´´s Hosp., Harvard Med. Sch., Boston, MA, USA
Abstract :
A custom built 512 channel system was designed to drive therapeutic ultrasound phased arrays. The system was used to drive a newly constructed 256 element phased array in a series of MR guided in vivo porcine experiments to coagulate deep seated tissue in the thigh, kidney, and liver. Thermal coagulation lesions greater than 5 cm3 in thigh and greater than 2 cm3 in the liver were generated using a single 20 second sonication. The feasibility of MR guided phased array ultrasound surgery to coagulate in vivo liver tissue in a large animal model has been shown
Keywords :
biomedical equipment; biomedical ultrasonics; coagulation; kidney; liver; surgery; ultrasonic arrays; 20 s; 256 element phased array; MR guided in vivo porcine experiments; deep seated tissue treatment; in vivo liver tissue coagulation; kidney; large animal model; large scale phased array; liver; medical instrumentation; sonication; therapeutic ultrasound phased arrays; thigh; Animals; Coagulation; In vivo; Large-scale systems; Lesions; Liver; Phased arrays; Surgery; Thigh; Ultrasonic imaging;
Conference_Titel :
Ultrasonics Symposium, 1998. Proceedings., 1998 IEEE
Conference_Location :
Sendai
Print_ISBN :
0-7803-4095-7
DOI :
10.1109/ULTSYM.1998.765241