DocumentCode
2812972
Title
Traceable calibration of Vector Signal Analyzers
Author
Humphreys, David A. ; Harper, Matthew R. ; Salter, Martin
Author_Institution
Nat. Phys. Lab., Teddington, UK
fYear
2010
fDate
28-28 May 2010
Firstpage
1
Lastpage
4
Abstract
We have developed a calibration technique for Vector Signal Analyzers (VSA) that provides traceability to primary standards. This technique has been implemented using standard commercial equipment. A multisine signal generator provides detailed magnitude and phase stimulus over the instrument bandwidth and a Digital Sampling Oscilloscope (DSO) measures the same multisine waveform. Traceability is provided via the DSO, calibrated against the primary standard (Electro-Optic Sampler). The VSA can be characterized over its full-bandwidth (magnitude and phase) at any operating frequency. The reproducibility of three different multi-sine measurements, averaged over an 80 MHz bandwidth was 3.4% (power) and 1.3 degrees (phase).
Keywords
calibration; electro-optical devices; microwave measurement; oscilloscopes; signal generators; spectral analysers; digital sampling oscilloscope; electro-optic sampler; multisine measurements; multisine signal generator; traceable calibration; vector signal analyzers; Bandwidth; Calibration; Frequency; Instruments; Oscilloscopes; Phase measurement; Sampling methods; Signal analysis; Signal generators; Standards development; digital sampling oscilloscopes; multisines; phase measurement; time-base correction; vector signal analyzer;
fLanguage
English
Publisher
ieee
Conference_Titel
Microwave Measurements Conference (ARFTG), 2010 75th ARFTG
Conference_Location
Anaheim, CA
Print_ISBN
978-1-4244-6364-0
Type
conf
DOI
10.1109/ARFTG.2010.5496315
Filename
5496315