    Assurance of Quality Control in Analytical Procedures Using Mass Spectrometry

  • Author

    Moldovan, Zaharie ; Schmutzer, Gabriella ; Alder, Alfredo

  • Author_Institution
    National Inst. of R&D for Isotopic & Molecular Technol., INCDTIM, Cluj-Napoca
  • Volume
    2
  • fYear
    2006
  • fDate
    25-28 May 2006
  • Firstpage
    205
  • Lastpage
    207
  • Abstract
    The aim of this paper was to obtain a practical procedure for processing the experimental data acquired by gas chromatography-mass spectrometry system to obtaining optimum precision. A number of more hundred compounds, belonging at different chemical class (including pharmaceutics) were detected as micro-pollutants in surface waters. They are in small concentrations ranging from 10 to 1000 of ng/1. The main objective of the papers is precise determination of this compounds using GC/MS area as experimental data. The calibration data are described by three type of function f1=ax+b, f2=ax and f3 =ax2+bx+c. The obtained precision is discussed in correlation with used function for two compounds, diazepam and triclosan
  • Keywords
    calibration; chromatography; mass spectroscopy; quality control; water pollution measurement; analytical procedures; gas chromatography-mass spectrometry; quality control assurance; surface water micropollutants; Calibration; Chemical industry; Environmentally friendly manufacturing techniques; Instruments; Linearity; Mass spectroscopy; Pharmaceuticals; Quality control; Temperature; Wastewater treatment;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Automation, Quality and Testing, Robotics, 2006 IEEE International Conference on
  • Conference_Location
    Cluj-Napoca
  • Print_ISBN
    1-4244-0360-X
  • Electronic_ISBN
    1-4244-0361-8
  • Type

    conf

  • DOI
    10.1109/AQTR.2006.254632
  • Filename
    4022955