    YAP-PET: a small animal positron emission tomograph based on YAP:Ce finger crystals

  • Author

    Guerra, A. Del ; Domenico, G. Di ; Scandola, M. ; Zavattini, G.

  • Author_Institution
    Dipt. di Fisica, INFN, Ferrara, Italy
  • Volume
    2
  • fYear
    1997
  • fDate
    9-15 Nov 1997
  • Firstpage
    1640
  • Abstract
    The authors have built a small animal positron emission tomograph (YAP-PET) based on four matrices of 400 YAP:Ce finger crystals (2×2×30 mm3 each) coupled to Hamamatsu R2486-06 position sensitive photomultipliers. The system consists of four detectors on a rotating gantry with a variable diameter between 10 and 25 cm. The read-out and data acquisition system are handled by off-shelf electronics and are controlled by in house developed programs. The tomograph has a total FOV of 4×4×4 cm3 that is appropriate for mice and rat studies. A spatial resolution ⩽2 mm over the useful FOV of 2×2×2 cm3 has been measured, with an average sensitivity of about 730 cps/μCi for a gantry diameter of 15 cm and an energy threshold of 50 keV
  • Keywords
    biological techniques; cerium; data acquisition; image resolution; photomultipliers; positron emission tomography; yttrium compounds; 10 to 25 cm; 50 keV; Hamamatsu R2486-06 position sensitive photomultipliers; YAP-PET; YAP:Ce; YAP:Ce finger crystals; YAlO3:Ce; YAlO3:Ce; data acquisition system; detectors; energy threshold; field of view; mice studies; rat studies; rotating gantry; sensitivity; small animal positron emission tomograph; spatial resolution; Animals; Control systems; Crystals; Data acquisition; Detectors; Electric variables control; Fingers; Mice; Photomultipliers; Radioactive decay;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1997. IEEE
  • Conference_Location
    Albuquerque, NM
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-4258-5
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1997.670632
  • Filename
    670632