• DocumentCode
    2836811
  • Title

    Simulation on the effects of dental implant threads to the osseointegration potential

  • Author

    Ahmad, Siti A. ; Kadir, Mohammed Rafiq Abdul ; Sulaiman, Eshamsul

  • Author_Institution
    Med. Implant Technol. Group, Univ. Teknol. Malaysia, Skudai, Malaysia
  • fYear
    2010
  • fDate
    Nov. 30 2010-Dec. 2 2010
  • Firstpage
    221
  • Lastpage
    226
  • Abstract
    In this paper, effect of the implant thread profile on the stability and osseointegration at the peri-implant bone were evaluated by finite element analysis. These study include five different common threads profile of implant body - buttress, reverse buttress, V-shape, sinusoidal and square - that were evaluated via finite element analysis. Three dimensional model of a mandible was reconstructed from CT dataset and a region of interest was selected at the left side around the first molar. The mandible was separated into two types - compact cortical and porous cancellous. This will allow for proper load transfer to the mandible during simulation of the mastication process. A generic abutment model was virtually connected to the implant body, and a crown representing the first molar was placed onto the abutment. The motion of the implant body relative to the bone was calculated under occlusal load ranging from 10N to 200N applied to the crown. Result showed that the reverse buttress and V-shape thread offer high stability by provide less micromotion at the dental implant body.
  • Keywords
    biomechanics; bone; dentistry; fasteners; finite element analysis; medical computing; physiological models; prosthetics; CT dataset; V-shape thread profile; buttress thread profile; compact cortical mandible; dental implant thread effects; finite element analysis; generic abutment model; implant body thread profile; implant osseointegration; implant stability; implant thread profile; load transfer; mandible 3D model; mastication simulation; osseointegration potential; periimplant bone; porous cancellous mandible; reverse buttress thread profile; sinusoidal thread profile; square thread profile; Bones; Dentistry; Finite element methods; Implants; Stability analysis; Stress; Teeth; dental implant; displacement; finite element analysis; micromotion; thread;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering and Sciences (IECBES), 2010 IEEE EMBS Conference on
  • Conference_Location
    Kuala Lumpur
  • Print_ISBN
    978-1-4244-7599-5
  • Type

    conf

  • DOI
    10.1109/IECBES.2010.5742232
  • Filename
    5742232