Development of a phantom to simulate a wide range of body sizes
Author
Lokitz, Stephen J. ; Wilson, Joshua M. ; Turkington, Timothy G.
Author_Institution
Duke Univ., Durham
Volume
5
fYear
2007
fDate
Oct. 26 2007-Nov. 3 2007
Firstpage
3713
Lastpage
3717
Abstract
X-ray image quality is highly dependent on object size. Diagnostic image quality can be maintained while controlling dose to a patient with tube current modulation schemes. For PET attenuation correction, much lower requirements are placed on CT image quality, but the minimum technique is still likely to be body size dependent. To investigate tube current modulation and non-diagnostic CT image quality effects on PET attenuation correction, and to aid in the development of new body-size dependent CT protocols for CT and PET/CT, we have developed a water-equivalent phantom that simulates a wide range of body sizes. Design criteria for the phantom include cross-sectional dimensions, shape, taper rate and material. Cross sectional dimensions and shape were desired to closely resemble different sized human torsos. Taper rate was important to achieve a smooth edge to reduce edge artifacts and to prevent partial volume effects. The phantom material needed to be easily machined, but still be tissue equivalent for CT imaging. The finished phantom was made of cast acrylic. Twelve one-inch plates of acrylic were machined into ovals with continuously tapered edges. After fabrication, the plates were stacked and secured together. The tolerance between plate diameters is less than 1.0 mm. The final dimensions are 12.3 x 22.4 cm at the small end and 38.8 x 49.0 cm at the large end. The phantom has an overall length of 29.0 cm and a mass of 23.2 kg. Initial results from this phantom include evaluation of image quality when using tube current modulation and evaluation of low-dose CT techniques for accuracy in PET attenuation correction.
Keywords
X-ray imaging; biomedical imaging; phantoms; positron emission tomography; CT imaging; PET attenuation correction; X-ray imaging; acrylic; body sizes; body-size dependent CT protocols; human torsos; phantom; Attenuation; Biological materials; Biological system modeling; Computed tomography; Image quality; Imaging phantoms; Positron emission tomography; Protocols; Shape; X-ray imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
Conference_Location
Honolulu, HI
ISSN
1095-7863
Print_ISBN
978-1-4244-0922-8
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2007.4436928
Filename
4436928
Link To Document