• DocumentCode
    2847057
  • Title

    Electrochemical sensor for the lead determination: development and characterisation

  • Author

    Loh, Han Chern ; Ahmad, Musa ; Taib, Mohd Nasir

  • Author_Institution
    Fac. of Eng. & Sci., Univ. Tunku Abdul Rahman, Selangor, Malaysia
  • fYear
    2005
  • fDate
    5-7 Sept. 2005
  • Firstpage
    118
  • Lastpage
    126
  • Abstract
    Lead electrochemistry sensor had been developed using screen printed electrode (SPE) technique with poly(vinyl) chloride (PVC) as printed matrix. The study has been carried out by using different type of electrodes. However, lead electrochemistry sensor had optimum response when deposited with mercury on the top of working electrode and in the buffer solution of pH 2. The optimum parameters for mercury deposition include deposited time = 15 s, deposited potential = 1000 mV, scanning range = 20 mV/s and amplitude pulse = 25. This sensor has linearity range of 1 - 50 ppm of Pb(II) R2 of 0.9997. The LOD of this sensor is lower than 1 ppm of Pb(II).
  • Keywords
    electrochemical sensors; electrodes; lead; polymers; 1000 mV; 15 s; Pb; electrochemical sensor; electrochemistry; lead determination; poly(vinyl) chloride; screen printed electrode; Contamination; Electrodes; Food technology; Humans; Lead compounds; Linearity; Mass spectroscopy; Sensor phenomena and characterization; Water pollution; Water resources; Electrochemistry Sensor; Lead; Mercury; Poly(vinyl) chloride; Screen Printed Electrode;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors and the International Conference on new Techniques in Pharmaceutical and Biomedical Research, 2005 Asian Conference on
  • Print_ISBN
    0-7803-9370-8
  • Type

    conf

  • DOI
    10.1109/ASENSE.2005.1564520
  • Filename
    1564520