• DocumentCode
    2853262
  • Title

    Verification of beam placement accuracy in gamma knife radiosurgery by a BANG3 polymer gel dosimeter

  • Author

    Watanabe, Yoichi ; Perera, Gerard M. ; Mooij, Rob B.

  • Author_Institution
    Dept. of Radiat. Oncology, Columbia Univ., New York, NY, USA
  • Volume
    4
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    2450
  • Abstract
    Polymer gel dosimeter BANG3(R) was used to measure beam placement accuracy of a Leksell Gamma Knife radiosurgery unit. The authors did two experiments over a five-month period. Both show a 1.8-mm (corresponding to the size of two pixels) shift of measured dose distributions relative to the isodose lines calculated by the treatment planning software in the left-right direction. There is no observable error in the anterior-posterior direction. The authors have considered several sources for the observed shift, which include phantom setup error, MRI image distortion due to field inhomogeneity and non-linearity, treatment planning software error, and design/manufacturing error of the radiosurgery unit. The observed error has not been identified using the quality assurance procedure recommended by the manufacturer. The authors have demonstrated, therefore, that the polymer gel dosimeter is useful to verify the beam placement accuracy of a Leksell Gamma Knife unit as a part of acceptance test of hardware and treatment planning software. They recommend that the treatment prescription should be made considering the accuracy limit presented in this study
  • Keywords
    dosimeters; gamma-ray applications; gels; polymers; quality control; radiation therapy; surgery; 1.8 mm; 5 month; BANG3 polymer gel dosimeter; accuracy limit; anterior-posterior direction; beam placement accuracy verification; gamma knife radiosurgery; left-right direction; measured dose distributions; quality assurance procedure; treatment planning software; treatment prescription; Distortion measurement; Imaging phantoms; Magnetic resonance imaging; Manufacturing; Polymer gels; Quality assurance; Size measurement; Software design; Software measurement; Software testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-6465-1
  • Type

    conf

  • DOI
    10.1109/IEMBS.2000.901294
  • Filename
    901294