Abstract
Proposes an algorithm based on simulated annealing for automatic seed matching and reconstruction using either 3 radiographic films or 3 fluoroscopic images. An algorithm for automated seed identification on fluoroscopic images is also proposed. The authors present results obtained for 112 clinical cases, 100 simulated cases and one phantom reconstruction with known seed locations. Results for clinical eases are compared with results obtained using the 3-film technique
Keywords
biological organs; diagnostic radiography; image matching; image reconstruction; medical image processing; radiation therapy; simulated annealing; 3-film technique; automated prostate implant seed localization; automated seed identification algorithm; clinical eases; fluoroscopic images; permanent implants; radiographic films; therapeutic nuclear medicine; Computed tomography; Computer vision; Cost function; Error correction; Image reconstruction; Imaging phantoms; Implants; Magnetic resonance imaging; Radiography; Simulated annealing;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
Conference_Location
Chicago, IL
ISSN
1094-687X
Print_ISBN
0-7803-6465-1
Type
conf
DOI
10.1109/IEMBS.2000.901440
Filename
901440
Link To Document