• DocumentCode
    2857648
  • Title

    Neutron measurements for intensity modulated radiation therapy

  • Author

    Ipe, N.E. ; Roesler, S. ; Jiang, S.B. ; Ma, C.-M.

  • Author_Institution
    Radiat. Phys. Dept., Stanford Linear Accel. Center, Menlo Park, CA, USA
  • Volume
    4
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    3234
  • Abstract
    The beam-on time for intensity modulated radiation therapy (IMRT) is increased significantly compared with conventional radiotherapy treatments. Further, the presence of beam modulation devices may potentially affect neutron production. Therefore, neutron measurements were performed for 15 MV photon beams on a Varian Clinac accelerator to determine the impact of IMRT on neutron dose equivalent to the patient
  • Keywords
    dosimetry; linear accelerators; neutron detection; radiation therapy; 15 MV; Varian Clinac accelerator; intensity modulated radiation therapy; neutron dose equivalent; neutron measurements; neutron production; Biomedical applications of radiation; Detectors; Gold; Intensity modulation; Neutrons; Optical modulation; Optical scattering; Particle beams; Performance evaluation; Production;