DocumentCode :
2860072
Title :
Use Case Demonstration: X-Ray/Ventilator
Author :
Arney, David ; Goldman, Julian M. ; Lee, Insup ; Llukacej, Ersel ; Whitehead, Susan
Author_Institution :
Pennsylvania Univ., Philadelphia
fYear :
2007
fDate :
25-27 June 2007
Firstpage :
160
Lastpage :
160
Abstract :
Patients who need an x-ray while under anesthesia are at risk because clinicians turn off the anesthesia ventilator for an x-ray and may forget to turn it back on afterwards. This problem occurred in a case study reported by the Anesthesia Patient Safety Foundation in 2004. We demonstrate a system in which automatic synchronization of the x-ray exposure with ventilation eliminates the need to turn off the ventilator to obtain an x-ray. This is accomplished by triggering the x-ray (simulated with a web camera) with the respiratory flow signal.
Keywords :
diagnostic radiography; Anesthesia Patient Safety Foundation; anesthesia ventilator; use case demonstration; x-ray exposure; x-ray/ventilator; Anesthesia; Context modeling; Control systems; Graphical user interfaces; Hospitals; Lungs; Medical simulation; Safety; Simple object access protocol; Software systems;
fLanguage :
English