• DocumentCode
    2867999
  • Title

    An 11k-Electrode 126-Channel High-Density Microelectrode Array to Interact with Electrogenic Cells

  • Author

    Frey, Urs ; Heer, Flavio ; Pedron, René ; Hafizovic, Sadik ; Greve, Frauke ; Sedivy, Jan ; Kirstein, Kay-Uwe ; Hierlemann, Andreas

  • Author_Institution
    ETH, Zurich
  • fYear
    2007
  • fDate
    11-15 Feb. 2007
  • Firstpage
    158
  • Lastpage
    593
  • Abstract
    A microelectrode array allows an arbitrary group of 126 electrodes to be selected from a total of 11,016 in order to do cell or neural recordings from areas of interest with 18 mum spatial resolution and 2.4 muv input-referred noise. Signals are amplified by 0 to 80dB, bandpass filtered (0.3 to 4kHz), and finally digitized (20kS/s, 8b). Example recordings from acute brain slices are shown
  • Keywords
    CMOS integrated circuits; brain; cellular biophysics; microelectrodes; neurophysiology; 0.18 micron; 0.3 to 4 kHz; acute brain slices; cell recording; electrogenic cells; high-density microelectrode array; neural recordings; Biosensors; Brain; Crosstalk; Data processing; Electrodes; Event detection; Field programmable gate arrays; Microelectrodes; Neurons; Solid state circuits;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Circuits Conference, 2007. ISSCC 2007. Digest of Technical Papers. IEEE International
  • Conference_Location
    San Francisco, CA
  • ISSN
    0193-6530
  • Print_ISBN
    1-4244-0853-9
  • Electronic_ISBN
    0193-6530
  • Type

    conf

  • DOI
    10.1109/ISSCC.2007.373636
  • Filename
    4242313