Abstract :
We report a method for improving the accuracy in determining blood flow rate and small vessel size by implementing a delay line filter when processing real-time spectral domain optical coherence tomography images of flow.
Keywords :
Doppler effect; biomedical optical imaging; blood flow measurement; blood vessels; delay filters; eye; medical image processing; optical delay lines; optical tomography; Doppler flow velocity; blood flow rate; blood vessel size; delay line filter; human retina; image processing; optical coherence tomography; spectral domain optical Doppler tomography; stationary scatterer filtering; Biomedical optical imaging; Delay lines; Filtering; High speed optical techniques; Optical filters; Optical imaging; Optical interferometry; Optical recording; Optical scattering; Tomography; (170.3340) Laser Doppler velocimetry; (170.3880) Medical and biological imaging; (170.4500) Optical coherence tomography;