25 megahertz gastrointestinal ultrasonography
Author
Wiersema, M.J. ; Reilly, C.R. ; Sanghvi, N.T. ; Hawes, R. ; Wiersema, L. ; Aust, C.
Author_Institution
Indianapolis Center for Adv. Res., Indiana Univ., IN, USA
fYear
1989
fDate
3-6 Oct 1989
Firstpage
845
Abstract
Endoscopic gastrointestinal ultrasonography provides the only method of in vivo evaluation of all gastrointestinal wall layers and surrounding structures. Higher frequencies (7.5 and 12 MHz) have been used due to the proximity of the transducer to the bowel wall. In an attempt to define the tissue characteristics at a macroscopic level and differentiate their structural content as a function of backscattering, the authors have employed a 25-MHz transceiver, high-speed digitizer, and high-resolution display in a conventional in vitro pulse-echo system. A linear B-mode scan constructed from digitized A-modes is stored in the microcomputer and displayed in gray scale (256 levels) on a high-resolution monitor. Transducer characterization yielded focal lateral and axial resolutions of 200 μm and 100 μm, respectively
Keywords
acoustic imaging; biomedical ultrasonics; patient diagnosis; 25 MHz; axial resolutions; backscattering; bowel wall; conventional in vitro pulse-echo system; digitized A-modes; endoscopic gastrointestinal ultrasonography; focal lateral resolution; gastrointestinal wall layers; gray scale; high-resolution display; high-resolution monitor; high-speed digitizer; in vivo evaluation; linear B-mode scan; microcomputer; structural content; surrounding structures; tissue characteristics; transceiver; Backscatter; Displays; Frequency; Gastrointestinal tract; In vitro; In vivo; Microcomputers; Transceivers; Transducers; Ultrasonography;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 1989. Proceedings., IEEE 1989
Conference_Location
Montreal, Que.
Type
conf
DOI
10.1109/ULTSYM.1989.67107
Filename
67107
Link To Document